Sysmex CA-1500全自动血凝仪PT-der法和Von-Clauss法检测纤维蛋白原的分析
引用本文:刘振杰,何燕香,徐宁,何文军,吴子安,黄丽英,李涛.Sysmex CA-1500全自动血凝仪PT-der法和Von-Clauss法检测纤维蛋白原的分析[J].国际检验医学杂志,2014,0(10):1333-1334,1370.
作者姓名:刘振杰  何燕香  徐宁  何文军  吴子安  黄丽英  李涛
作者单位:刘振杰 (广东省中医院检验科,广东广州,510370); 何燕香 (广东省中医院检验科,广东广州,510370); 徐宁 (广东省中医院检验科,广东广州,510370); 何文军 (广东省中医院检验科,广东广州,510370); 吴子安 (广东省中医院检验科,广东广州,510370); 黄丽英 (广东省中医院检验科,广东广州,510370); 李涛 (广东省中医院检验科,广东广州,510370);
摘    要:目的评价Sysmex CA-1500全自动血凝仪PT-der法和Von-Clauss法测定血浆纤维蛋白原(Fib)的相关性和准确性。方法采用Sysmex CA-1500全自动血凝仪PT-der法和Von-Clauss法分别测定755例血液标本的血浆Fib浓度;高Fib浓度标本的稀释比例为1∶8、2∶7、3∶6、4∶5、5∶4、6∶3,以稀释比例为横坐标,以2种方法测得的Fib浓度为纵坐标,进行简单线性回归分析。结果 Fib:2.0~〈6.0g/L时,Von-Clauss法检测Fib值高于PT-der法(P〈0.01);Fib:〈2.0g/L,〉6.0g/L时,PT-der法检测Fib值高于Von-Clauss法(P〈0.01)。PT-der法和Von-Clauss法检测的线性回归方程分别为:Y=4.537 7 X+1.551 3(R2=0.897 3),Y=7.792 2 X+0.290 0(R2=0.980 5)。结论 Von-Clauss法能较好地反映人体内具有凝血功能的Fib水平。

关 键 词:纤维蛋白原  Von-Clauss法  PT-der法  线性回归

Analysis on fibrinogen detection by using PT-der assay and Von-Clauss assay on Sysmex CA-1500 Automated Coagulation Analyzer
Liu Zhenjie,He Yanhong,Xu Ning,He Wenjun,Wu Zi′an,Huang Liying,Li Tao.Analysis on fibrinogen detection by using PT-der assay and Von-Clauss assay on Sysmex CA-1500 Automated Coagulation Analyzer[J].International Journal of Laboratory Medicine,2014,0(10):1333-1334,1370.
Authors:Liu Zhenjie  He Yanhong  Xu Ning  He Wenjun  Wu Zi′an  Huang Liying  Li Tao
Institution:Liu Zhenjie, He Yanhong, Xu Ning, He Wenjun, Wu Zi′an, Huang Liying, Li Tao
Abstract:Objective To evaluate the relevance and accuracy of fibrinogen (Fib) detection by using PT-der assay and Von-Clauss assay on Sysmex CA-1500 Automated Coagulation Analyzer .Methods PT-der assay and Von-Clauss assay on Sysmex CA-1500 Automated Coagulation Analyzer were employed to detect the plasma Fib concentrations of 755 blood samples .The dilution ratios of samples with high Fib concentration were 1∶8 ,2∶7 ,3∶6 ,4∶5 ,5∶4 ,6∶3 ,respectively .The dilution ratios were served as the abscissa ,and the Fib concentrations measured by two methods as the ordinate ,a simple linear regression analysis was per-formed .Results When Fib concentration was in 2 .0- <6 .0 g/L ,the Fib value obtained by Von-Clauss assay was higher than that by PT-der assay(P〈0 .01) .When Fib concentration was below 2 .0 g/L or above 6 .0 g/L ,the Fib value obtained by PT-der assay was higher than that by Von-Clauss assay(P〈0 .01) .The linear regression equations of PT-der assay and Von-Clauss assay were Y=4 .537 7X+1 .551 3(R2 =0 .897 3) ,Y = 7 .792 2X+ 0 .290 0(R2 =0 .980 5) ,respectively .Conclusion Von-Clauss assay can better reflect the Fib level of human body which has a blood clotting function .
Keywords:fibrinogen  Von-Clauss assay  PT-der assay  linear regression