FloTrac/Vigileo系统用于冠状动脉搭桥术中血流动力学监测
引用本文:卢家凯,朱琛,王义军,卿恩明.FloTrac/Vigileo系统用于冠状动脉搭桥术中血流动力学监测[J].心肺血管病杂志,2012,31(2):101-105.
作者姓名:卢家凯  朱琛  王义军  卿恩明
作者单位:1. 100029,北京 首都医科大学附属北京安贞医院-北京市心肺血管疾病研究所麻醉科
2. 北京中国煤炭总医院
摘    要:目的:观察FloTrac/Vigileo系统在非体外循环冠状动脉旁路移植术(OPCABG)血流动力学监测中应用的可行性,分析其与肺动脉导管(PAC)心排量监测(CCO)结果的相关性。方法:随机选择50例择期OPCABG的患者,所有患者同时应用连续心排出量(APCO)和CCO进行心排量(CO)监测,分别在切皮前、吻合左冠状动前降支、吻合回旋支、吻合后降支、关胸后5个时间点,测定2种方法监测的CO、心排指数(CI)及其他血流动力学参数。结果:术中在固定器放置、改变心脏位置、血管活性药物及使用正性肌力药情况下,APCO监测结果均有所改变,变化趋势与PAC监测结果一致。2种监测方法在各时间点监测的CO值的相关系数(r)分别为0.70,0.59,0.78,0.74及0.85,各时间点CI差值的偏倚范围分别为:(0.39±0.06)L.min-1.m-2、(0.48±0.12)L.min-1.m-2、(0.26±0.06)L.min-1.m-2、(0.27±0.06)L.min-1.m-2及(0.30±0.05)L.min-1.m-2,提示2种方法监测结果有较好的相关性。2种监测方法所得(SVR)在吻合血管时变化趋势与各自监测所得CO变化趋势相反。结论:APCO用于OP-CABG术中连续CO监测具有可行性,与PAC监测法所得CO结果相关,监测结果体现了较好的即时性和动态性。

关 键 词:FloTrac/Vigileo系统  肺动脉导管  心排量监测  冠状动脉旁路移植术

Application of arterial pressure-based cardiac output measurement in patients undergoing off-pump coronary artery bypass surgery
LU Jiakai , ZHU Chen , WANG Yijun , Qing Enming.Application of arterial pressure-based cardiac output measurement in patients undergoing off-pump coronary artery bypass surgery[J].Journal of Cardiovascular and Pulmonary Diseases,2012,31(2):101-105.
Authors:LU Jiakai  ZHU Chen  WANG Yijun  Qing Enming
Institution:Department of Anesthesiology.Capital Medical University affiliated Beijing Anzhen Hospital,Beijing Institute of Heart,Lung and Blood Vessel Diseases,Beijing,100029,China
Abstract:Objective:To observe the feasibility of FloTrac/Vigileo system for cardiac output measurement and to evaluate the associativity of monitoring results between FloTrac/Vigileo system compared and pulmonary artery catheter(PAC) for cardiac output measurement in patients undergoing off-pump coronary atery bypass grafting(OPCABG).Methods:Fifty patients undergoing off-pump coronary artery bypass grafting were cruited randomly.All patients accepted cardiac output monitiring by PAC and APCO FloTrac/Vigileo system at the same time.Following data were designed to be monitored including continuous cardiac output(CCO),cardiac index(CI) and other hemodynamic parameters at 5 points: after anaesthesia induction(T1),anastomosis of left internal mammary artery to left anterior descending artery(T2),left circumflex anastomosis(T3),posterior descending artery anastomosis(T4),after sternal closure(T5).Results: Correlation index of both monitoring methods of CCO measurements were 0.70,0.59,0.78,0.74 and 0.85 at each time point.The bias range of CI difference were(0.39±0.06) L·min-1·m-2,(0.48±0.12) L·min-1·m-2,(0.26±0.06) L·min-1·m-2,(0.27±0.06) L·min-1·m-2 and(0.30±0.05) L·min-1·m-2 at each time point.The variation trend of SVR and CO derived from both methods were oppsite when different coronary artery anastomosis.Conclusion: APCO reflects timely,accurate and continuous hemodynamic changes in patients undergoing OPCABG.The monitoring results derived from APCO FloTrac/Vigileo system and PAC are well correlated in patients undergoing OPCABG.
Keywords:FloTrac/Vigileo system  Pulmonary artery catheter  Cardiac output measurement  Coronary artery bypass grafting