3种仪器测量中央角膜厚度的比较研究
引用本文:成拾明,周霞,许玲. 3种仪器测量中央角膜厚度的比较研究[J]. 现代中西医结合杂志, 2012, 21(18): 1943-1945
作者姓名:成拾明  周霞  许玲
作者单位:湖北医药学院附属太和医院,湖北,十堰,442000
摘    要:目的比较Pentacam、SP 3000和A型超声测量正常人中央角膜厚度的差异性及一致性。方法分别用Pentacam、SP 3000和A型超声测量40例正常人右眼的中央角膜厚度。结果 Pentacam、SP 3000和A超测量的中央角膜厚度分别为(535.8±33.8)μm、(508.6±35.5)μm和(530.7±36.9)μm;A型超声与Pentacam测量的中央角膜厚度无显著性差异(P=0.517),两者的差值为(5.4±7.9)μm,两者95%的一致性区间为(-18.1~7.8)μm;A型超声与SP3000测量的中央角膜厚度有显著性差异(P=0.006),两者的差值为(21.2±7.9)μm,95%一致性区间为(8.6~35.7)μm;SP 3000与Pentacam测量结果有显著性差异(P=0.001),两者的差值为(27.3±7.9)μm,两者95%的一致性区间为(13.7~40.8)μm。结论 Pentacam与A型超声测量正常人中央角膜厚度的差值较小,一致性较好,在测量正常人中央角膜厚度时可以相互替代使用;Pentacam与SP 3000、A型超声与SP 3000测量中央角膜厚度的一致性较差,在临床应用中要引起注意。

关 键 词:Pentacam  SP3000  A型超声  中央角膜厚度

Comparison of central corneal thickness measurements by three different instruments
Cheng Shiming , Zhou Xia , Xu Ling. Comparison of central corneal thickness measurements by three different instruments[J]. Modern Journal of Integrated Chinese Traditional and Western Medicine, 2012, 21(18): 1943-1945
Authors:Cheng Shiming    Zhou Xia    Xu Ling
Affiliation:(Taihe Hospital Affiliated to Hubei Medical College,Shiyan 442000,Hubei,China)
Abstract:Objective It is to compare the difference and agreement in central corneal thickness measurements by Pentacam,SP 3000 and A-scan.Methods Totally 40 eyes of 40 volunteers were measured with Pentacam,SP 3000 and A-scan in central corneal thickness.Results The measurements of central corneal thickness were(535.8±33.8)μm,(508.6±35.5) μm and(530.7±36.9)μm respectively with Pentacam,SP 3000 and A-sca.The difference between Pentacam and A-scan was(5.4±7.9)μm,no statistically significant difference could be found(P=0.517),the 95% limit of agreement was(-18.1~7.8)μm.There was statistically significant difference between SP 3000 and A-scan measurements(mean difference was(21.2±7.9)μm,P=0.006),the 95% limit of agreement was(8.6~35.7)μm.There was statistically significant difference between SP 3000 and A-scan measurements(mean difference was(27.3±7.9)μm,P=0.001),the 95% limit of agreement was(13.7~40.8)μm.Conclusion Little difference and good agreement can be found between Pentacam and A-scan in central corneal thickness measurements.They can be used interchangeably in central corneal thickness of normal people.The agreement between Pentacam and SP 3000,A-scan and SP 3000 are bad.Attention should be paid to the differences between each two in the measurements of central corneal thickness.
Keywords:Pentacam  SP 3000  A-scan  central corneal thickness