Stanford B型主动脉夹层发病24h内腔内修复术30例
引用本文:汤敬东,黄俊峰,左克强,钱振宇,杭闻曌,龚凯,黄道景,吴春红,张文艳,杨明烽,景在平,王玉琦.Stanford B型主动脉夹层发病24h内腔内修复术30例[J].中华普通外科杂志,2010,25(3).
作者姓名:汤敬东  黄俊峰  左克强  钱振宇  杭闻曌  龚凯  黄道景  吴春红  张文艳  杨明烽  景在平  王玉琦
作者单位:1. 上海同济大学附属同济医院血管外科,200065
2. 上海,第二军医大学附属长海医院
3. 上海,复旦大学附属中山医院血管外科
摘    要:目的 探讨Stanford B型主动脉夹层(aortic dissection,AD)发病24h内腔内修复术的疗效.方法 回顾性分析2007年6月至2008年1O月急诊行腔内修复Stanford B型主动脉夹层30例临床资料.本组腔内修复术均在胸痛症状出现后2.4h内完成;采用全身麻醉,在x线透视下将覆膜支架经股动脉放置在夹层第l裂口位置;其中1例第1裂口位于腹腔干对侧缘,运用开窗支架技术实施腔内修复术.术后1周、1、3、6、12个月行螺旋CT检查,观察手术疗效以及有无内漏、支架移位和内脏血供等.结果 30例支架均成功释放,无支架移位,无中转开胸手术,即刻造影I型内漏4例(13.4%),近端球囊扩张后3例消失,1例小于10%;所有患者真腔供血均得到改善.术后30 d死亡1例(3.3%),可能缘于术后夹层破裂;30例术后随访1-19个月,平均(12 4-8)个月,术后6个月发现1例(3.3%)I型内漏;无其他新发内漏出现.随访期间死亡1例(3.3%),该患者术后2月出现急性肝功能衰竭,死因不明.结论 24 h内腔内修复Stanford B型主动脉夹层可明显降低急性起病期患者死亡率,但远期效果还有待进一步观察.

关 键 词:动脉瘤  夹层  支架  急性期

Endovascular stent-grafts for acute Stanford type B aortic dissection
TANG Jing-dong,HUANG Jun-feng,ZUO Ke-qiang,QIAN Zhen-yu,HANG Wen-zhao,GONG Kai,HUANG Dao-jing,WU Chun-hong,ZHANG Wen-yan,YANG Ming-feng,JING Zai-ping,WANG Yu-qi.Endovascular stent-grafts for acute Stanford type B aortic dissection[J].Chinese Journal of General Surgery,2010,25(3).
Authors:TANG Jing-dong  HUANG Jun-feng  ZUO Ke-qiang  QIAN Zhen-yu  HANG Wen-zhao  GONG Kai  HUANG Dao-jing  WU Chun-hong  ZHANG Wen-yan  YANG Ming-feng  JING Zai-ping  WANG Yu-qi
Abstract:Objective To assess the results of emergency endovascular stent-grafting for patients with acute Stanford type B aortic dissection(type B AAD)within 24 hours of onset.Methods Between June 2007 and October 2008,30 patients with acute type B aortic dissection underwent emergency endOVascular stent-grafting within 24 hours of presentation.Under general anesthesia,stent-graft was deploved at the proper position of first tear entry through femoral artery under X-ray monitering.Follow-up by CT was performed 1 w,1 m ,3 m,6 m,1 y postoperatively to observe the efficacy and complications such as endoleak,migration and fracture of stent-graft. Result The technical success rate was 100%;13.4%(4 cases) endoleak rate was identified immediately after deployment.Follow up was made between 1 month to 19 months,averaging at(12±8)months,3.3%type-1 endoleak Was observed after 6 months;One patient died within 30 days possibly of dissection rupture;One patient died of acute liver failure during the follow-up. Conclusion Endovascular repair with stent-graft within 24 hours of presentation was effectivefor the treatment of acute type B aortic dissection.
Keywords:Aneurysm  dissecting  Stents  Acute phase