腺苷脱氨酶自建检测系统及评价
引用本文:汪小葛,唐新,浦雄勇,孙王伟. 腺苷脱氨酶自建检测系统及评价[J]. 检验医学与临床, 2013, 10(2): 169-170
作者姓名:汪小葛  唐新  浦雄勇  孙王伟
作者单位:江苏省昆山市第一人民医院检验科,215300
摘    要:目的比较北京利德曼生化股份有限公司生产的腺苷脱氨酶(ADA)检测试剂盒在Roche ModolarDPP全自动生化分析仪上建立的检测系统与Roche Modolar DPP可溯源的参考系统血清ADA检测结果之间的可比性和临床上可接受性。方法依据临床和实验室标准化协会(CLSI)标准中EP9-A文件对自建检测系统与可溯源参考系统检测门诊患者的40份新鲜血清ADA含量进行线性回归分析。结果自建系统和参考系统血清ADA检测结果具有很好的可比性。回归方程为Y=0.998 7 X+0.090 2,其相关系数r2=0.998 4。结论自建系统血清ADA检测结果与Roche Modolar DPP可溯源的标准参考检测系统具有很好的可比性,可以代替参考系统检测临床标本,能够被临床所接受。

关 键 词:腺苷脱氨酶  检测系统  线性回归分析

Comparison and evaluation of the self-established assay system for adenosine deaminase
WANG Xiao-ge,TANG Xin,PU Xiong-yong,SUN Wang-wei. Comparison and evaluation of the self-established assay system for adenosine deaminase[J]. Laboratory Medicine and Clinic, 2013, 10(2): 169-170
Authors:WANG Xiao-ge  TANG Xin  PU Xiong-yong  SUN Wang-wei
Affiliation:(Department of Laboratory Medicine,Kunshan First People′s Hospital,Kunshan,Jiangsu 215300,China)
Abstract:Objective To evaluate the comparability and clinical acceptability of the results of serum adenosine deaminase(ADA) measured by self-established assay system with Beijing Leadman ADA kits in Roche Modular DPP clinical chemistry analyzers compared with traceable Roche Modular P reference assay system.Methods The concentrations of ADA in 40 fresh outpatient serum samples were analyzed in duplicate with two self-established systems and traceable Roche Modular reference system.The linear regression analysis was calculated according to National Committee for Clinical Laboratory Standards CLSI document EP9-A files.Results The comparability of serum ADA measurement results by self-established systems compared with Roche Modular reference system was acceptable.The regression equation was Y=0.998 7X+0.090 2,and the correlation coefficient(r2) was 0.998 4.Conclusion The results of serum ADA measured by our self-established systems have good comparability compared with Roche Modular P reference system,and it can replace the reference system to detect clinical specimens,which could be acceptable by clinical application.
Keywords:adenosine deaminase  assay system  linear regression analysis