不同长度齿突对螺钉把持力的测试
引用本文:方弘伟,王向阳,施达,陈卓,赵云珍,王晓成,黄晓东,欧阳钧,吴增晖. 不同长度齿突对螺钉把持力的测试[J]. 医用生物力学, 2009, 24(3): 189-192
作者姓名:方弘伟  王向阳  施达  陈卓  赵云珍  王晓成  黄晓东  欧阳钧  吴增晖
作者单位:金华市中医院 骨二科;温州医学院附属二院 脊柱外科;金华职业技术学院 统计学教研室;金华市中医院 骨二科;金华市中医院 骨二科;金华市中医院 骨二科;金华市中医院 骨二科;南方医科大学 解剖教研室;广州军区总医院
摘    要:目的 评价不同长度齿突骨折螺钉内固定的生物力学效果。方法 收集20具新鲜成人枢椎标本,按齿突螺钉固定技术钻孔形成针道,模拟齿突II型骨折,测量齿突端针道长度,置入螺钉后测试旋入力矩、最大轴向拨出力和刚度测试,解剖观察钉道骨质构成。结果 钉道长度8.3mm~12.3mm,钉道长度与螺钉最大拔出力相关回归分析:相关系数r=0.28, 回归方程式:Y=156.69+67.97X (Y::钉道长 X:拔出力),大于10mm和小于10mm两组间旋入力矩、最大轴向拨出力和刚度t检验无统计学上差异(P>0.05),齿尖部由皮质骨构成。结论 齿突长度和骨折端的稳定度关系不密切、与最大轴向拨出力成正比关系但并不显著,短于1cm的齿突对螺钉仍有较好的把持力。

关 键 词:齿突骨折;内固定;螺钉;生物力学;骨折固定
收稿时间:2008-08-04
修稿时间:2008-11-21

Biomechanical evaluation of anterior screw fixation for type II fractures of the odontoid process: The influence of different lengths of the screws
FANG Hong-wei,WANG Xiang-yang,SHI D,CHEN Zhuo,ZHAO Yun-zhen,WANG Xiao-cheng,HUANG Xiao-dong,OUYANG Jun and WU Zheng-hui. Biomechanical evaluation of anterior screw fixation for type II fractures of the odontoid process: The influence of different lengths of the screws[J]. Journal of Medical Biomechanics, 2009, 24(3): 189-192
Authors:FANG Hong-wei  WANG Xiang-yang  SHI D  CHEN Zhuo  ZHAO Yun-zhen  WANG Xiao-cheng  HUANG Xiao-dong  OUYANG Jun  WU Zheng-hui
Affiliation:Department of Orthopaedic surgery, Traditional chinese medical hospital of Jinhua;Second Affiliated Hospital of Wenzhou Medical College;College of Proffession and Technique;Department of Orthopaedic surgery, Traditional chinese medical hospital of Jinhua;Department of Orthopaedic surgery, Traditional chinese medical hospital of Jinhua;Department of Orthopaedic surgery, Traditional chinese medical hospital of Jinhua;Department of Orthopaedic surgery, Traditional chinese medical hospital of Jinhua;Department of Anatomy, Southern Medical University;Guangzhou General Hospital of Army
Abstract:Objectives To determine the influence of different lengths of the anterior odontoid screws on the fixation stability. Methods Twenty cadaveric C2 vertebrae were harvested and an osteotomy was performed to simulate a Type II fracture pattern in each vertebra. The angle of type II fracture line was 0°in 6 specimens, 17°in 8, and 25°in 6. In this way, Different lengths of fractured odontoid processes were created and then divided into 2 groups (<10mm vs >10mm). The fractures were then treated by anterior screw fixation. And insertional torque, and pull-out force and stiffness of the bone-screw was tested. Results The lengths of the screws bedded in fractured odontoid process were 8.3mm-12.3mm. The relationship between the pull-out strength (Y) and the length of the screw (X) was determined by the regression equation (Y=156.69+67.97X, r=0.28). However, No significant differences of the insertional torque, and pull-out force and stiffness of the bone-screw were found between the <10mm and >10mm groups. Conclusions Bone-screw stability is not closely related to the distal part of the fractured odontoid process. Anterior screw fixation could be used even when distal fractured part of the odontoid process less than 10mm.
Keywords:Odontoid fracture   Internal fixation   Bone screw   Biomechanics