首页 | 本学科首页   官方微博 | 高级检索  
     


Targeting for thalamic deep brain stimulator implantation without computer guidance: assessment of targeting accuracy
Authors:Alterman R L  Reiter G T  Shils J  Skolnick B  Arle J E  Lesutis M  Simuni T  Colcher A  Stern M  Hurtig H
Affiliation:Institute for Neurology and Neurosurgery, Beth Israel Medical Center, New York, NY 10128, USA. ralterma@bethisraelny.org
Abstract:The authors assess the accuracy of targeting nucleus ventralis intermedius (Vim) with fast spin echo inversion recovery (FSE/IR) magnetic resonance imaging (MRI) in 18 successful deep brain stimulator (DBS) implants for medically refractory tremor. FSE/IR-MRI-derived coordinates are compared to the final coordinates employed for DBS lead placement, selected with intraoperative neurophysiology. The authors conclude that FSE/IR MRI is sufficiently reliable to serve as the sole means of anatomically targeting Vim for DBS lead placement. An independent computer workstation is not required for accurate targeting; however, intraoperative neurophysiology remains essential.
Keywords:
本文献已被 PubMed 等数据库收录!