Incomplete skin representation in digital mammograms |
| |
Authors: | Burgess Arthur E Kang Haejin |
| |
Affiliation: | Radiology Department, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts 02115, USA. burgess@bwh.harvard.edu |
| |
Abstract: | Flat-panel digital detector systems have limited dynamic range and saturate at a particular x-ray exposure. Hence some of the breast edge may not be represented in the displayed image. We developed a model to estimate the amount of skin loss. Model predictions agreed well with phantom measurements. In our database of 884 clinical digital mammograms, 98% had saturated backgrounds. The estimated skin loss exceeded 0.5 mm in 5% of images and 1.0 mm in 0.7% of images. Any skin thickening that is present should still be visualized, so we conclude that any skin-line loss may not be of clinical significance. |
| |