首页 | 本学科首页   官方微博 | 高级检索  
     


Flexor tendon repairs: the impact of fiberwire on grasping and locking core sutures
Authors:Miller Brian  Dodds Seth D  deMars Adam  Zagoreas Niko  Waitayawinyu Thanapong  Trumble Thomas E
Affiliation:Biomechanics Laboratory, Harborview Medical Center, Department of Orthopaedics and Sport Medicine, University of Washington, Seattle, WA 98195, USA.
Abstract:PURPOSE: Immediate surgical repair and early mobilization are essential in preventing adhesion formation and finger stiffness. A new polyethylene-based, braided suture material, Fiberwire (Arthrex, Naples, FL), touting increased strength, presents the potential for stronger repairs and, therefore, earlier active motion after surgery with a greater safety margin. The purpose of this biomechanic study was to investigate the differences in gap formation, tensile strength, and mode of failure for 2 distinct repair techniques using nylon, Ethibond (Ethicon, Somerville, NJ), and Fiberwire. METHODS: Human cadaver flexor tendons were harvested and repaired in a randomized fashion with either the Strickland or Massachusetts General Hospital (MGH) repairs using either nylon, Ethibond, or Fiberwire. Twelve tendons per group were repaired for each combination of material and method. During load-to-failure testing, 2-mm gap force and maximum tensile strength were statistically analyzed. RESULTS: Strickland repairs failed by suture pull-out in 74% of repairs, whereas 99% of the MGH repairs failed by suture breakage. For MGH repairs, Fiberwire suture provided significantly more tensile strength than Ethibond and nylon. For Strickland repairs, where the mode of failure was more often by suture pull-out rather than breakage, differences between type of suture were not significant. When comparing repair techniques using Fiberwire, the MGH repair was significantly stronger than the Strickland repair. CONCLUSIONS: Biomechanic testing shows that Fiberwire outperforms both Ethibond and nylon suture when using a locked flexor tendon repair suture (MGH repair) but not when using a grasping-type, nonlocking repair (Strickland repair).
Keywords:Fiberwire suture   flexor tendon   repair   biomechanics   tendon laceration   zone II
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号