首页 | 本学科首页   官方微博 | 高级检索  
检索        


Evaluation of a commercialized in situ hybridization assay for detecting human papillomavirus DNA in tissue specimens from patients with cervical intraepithelial neoplasia and cervical carcinoma
Authors:Guo Ming  Gong Yun  Deavers Michael  Silva Elvio G  Jan Yee Jee  Cogdell David E  Luthra Rajyalashmi  Lin E  Lai Hung Cheng  Zhang Wei  Sneige Nour
Institution:Department of Pathology, Unit 58, The University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030-4009, USA. mguo@mdanderson.org
Abstract:To evaluate a commercialized in situ hybridization (ISH) assay for detecting human papillomavirus (HPV) DNA, we compared the ability of a new ISH probe, Inform HPV III (Ventana Medical Systems, Tucson, AZ), to that of PCR assays to detect HPV DNA in cervical tissue specimens with normal cervix (20 cases), cervical intraepithelial neoplasia (CIN; CIN 1, 27 cases; CIN 2, 28 cases; and CIN 3, 33 cases), and cervical carcinoma (29 cases). General HPV DNA was detected using consensus primer-mediated PCR assays. HPV genotyping was performed by using EasyChip HPV blot (King Car Yuan Shan Institute, I-Lan, Taiwan). HPV16 integration status (E2/E6 ratio) was determined by using quantitative real-time PCR. Our findings showed that the ISH and PCR had fair to good agreements in detecting HPV DNA across all CIN categories without significant differences (Kappa coefficient, 0.34 to 0.63; P = 0.13 to 1.0). However, ISH detected significantly fewer HPV-positive cases in carcinoma than PCR did (Kappa coefficient, 0.2; P = 0.03). Eleven cases with ISH PCR+ results had HPV types that can be detected by Inform HPV III. Five carcinoma cases with ISH PCR+ results showed a significantly higher level of integrated HPV16 (P = 0.008) than did the ISH+ cases. As a consequence, lower copy numbers of episomal HPV16 in carcinoma might be the cause for the false-negative ISH results. Although the punctate signal pattern of HPV significantly increased with the severity of disease (P trend = 0.01), no significant difference in the HPV16 integration status was observed between the cases with a punctate signal only and the cases with mixed punctate and diffuse signals (P = 0.4). In conclusion, ISH using the Inform HPV III probe seems comparable to PCR for detecting HPV DNA in cervical tissue with CINs. False-negative ISH results appear to be associated with the lower copy numbers of the episomal HPV16 but not with the ability of the Inform HPV III probe to detect specific HPV types. In addition, signal patterns, especially a mixed punctate and diffuse pattern of HPV, cannot be reliably used to predict viral integration status.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号