尿干化学与流式细胞术联合用于尿液有形成分镜检筛选的研究与应用
引用本文:马骏龙,丛玉隆,陆玉静,王成彬,刘培培,张永强,凌励.尿干化学与流式细胞术联合用于尿液有形成分镜检筛选的研究与应用[J].中华检验医学杂志,2011,34(6):494-500.
作者姓名:马骏龙  丛玉隆  陆玉静  王成彬  刘培培  张永强  凌励
作者单位:1. 解放军总医院检验科,北京,100853
2. 希森美康医用电子(上海)有限公司
摘    要:目的 制定针对UF-1000i尿液分析流水线(由UF-1000i尿流式分析仪和AX-4030尿干化学分析仪组成)自动化检测结果的尿液镜检复检规则.方法 收集2009年9月至2010年2月解放军总医院尿液常规标本共2 839份.首先利用UF-1000i尿液分析流水线对2 839份尿液标本进行有形成分分析(包括RBC、WB...

关 键 词:尿分析  化学  分析  显微镜检查  自动分析

Study and application of the combination of urine dry chemistry examination and urine sediment flow cytometry examination on the microscopic screening of the urine formed component
MA Jun-long,CONG Yu-long,LU Yu-jing,WANG Cheng-bin,LIU Pei-pei,ZHANG Yong-qiang,LING Li.Study and application of the combination of urine dry chemistry examination and urine sediment flow cytometry examination on the microscopic screening of the urine formed component[J].Chinese Journal of Laboratory Medicine,2011,34(6):494-500.
Authors:MA Jun-long  CONG Yu-long  LU Yu-jing  WANG Cheng-bin  LIU Pei-pei  ZHANG Yong-qiang  LING Li
Abstract:Objective To establish the proper review rules for the microscopic screening of urine analyzed by UF-1000i automatic urinalysis work station (composed of UF-1000i urine flow cytometer and AX-4030 urine dry chemical analyzer).Methods A total of 2 839 random urine samples were collected at Chinese People′s Liberation Army General Hospital from September 2009 to February 2010, and were analyzed using UF-1000i urinalysis work station.The parameters obtained from UF-1000i and AX-4030 included RBC, WBC, CAST and ERY, LEU, PRO.After analysis by urinalysis work station, each urine sample was examined microscopically by two technologists using double-blind method.The average results got from the two technologists were regarded as the judging criterion.Based on the criterion, the review rules for the 2 839 urine samples tested by urinalysis work station were created and adjusted, and the true positive rate, false positive rate, true negative rate, false negative rate and review rate of these review rules were calculated.After that, 299 randomly selected urine samples were tested to validate these review rules.Omission diagnostic rate and review rate were used to assess the clinical practicability of the review rules.Results Thirty seven rules for microscopic review and twenty seven rules without further microscopic examination were set up based on six parameters using UriAccess 3.0 Software.The microscopic examination result was taken as the judging criterion, the consistency rate of these rules was 81.11%(2 311/2 839), the true positive rate was 40.51%(1 150/2 839), the false positive rate was 16.17%(459/2 839), the true negative rate was 41.00%(1 164/2 839), the false negative rate(omission diagnostic rate) was 2.43%(69/2 839) and the review rate was 18.28% (519/2 839).Additional 299 urine samples were assayed using UriAccess3.0 software to further verify these review rules.The consistency rate was 82.27%(246/299), the true positive rate was 36.12%(108/299), the false positive rate was 16.39%(49/299), the true negative rate was 46.15%(138/299), the false negative rate(omission diagnostic rate) was 1.34%(4/299), the review rate was 19.06%(57/299). The 4 false negative samples selected by these review rules did not come from the nephropathy department or the urology department.Microscopic results of RBC and WBC form these 4 samples ranged 3-8 cells/HP. Thus, these review rules could avoid the missed diagnosis of those patients with severe renal dysfunction.Conclusion The review rules established from this study for the UF-1000i urinalysis work station can effectively detect abnormal urine samples and improve the efficiency and the quality of urinalysis in routine clinical practice.
Keywords:Urinalysis  Chemistry  analytical  Microscopy  Autoanalysis