首页 | 本学科首页   官方微博 | 高级检索  
     


Pulmonary nodules: preliminary experience with three-dimensional evaluation
Authors:Revel Marie-Pierre  Lefort Catherine  Bissery Alvine  Bienvenu Marie  Aycard Laetitia  Chatellier Gilles  Frija Guy
Affiliation:Department of Radiology, Assistance Publique des H?pitaux de Paris/INSERM, Georges Pompidou European University Hospital, 20 Rue Leblanc, 75015 Paris, France. marie-pierre.revel@hop.egp.ap-hop-paris.fr
Abstract:PURPOSE: To evaluate software designed to calculate pulmonary nodule volume in three dimensions. MATERIALS AND METHODS: Fifty-four solid noncalcified pulmonary nodules measuring 5-18 mm in diameter were studied with computed tomographic (CT) volumetric software. Baseline CT examinations were performed for various indications by using four-detector row multisection CT units, 1.25- or 2.50-mm sections, and a standard reconstruction algorithm. The percentage of successful nodule segmentations, as well as intraobserver variability, interreader agreement, and global repeatability of calculated volumes, was determined on the basis of consecutive measurements performed three times by three different radiologists by using the Bland and Altman method. The software was used to calculate the doubling time of 22 nodules for which a final diagnosis and comparable CT scans were available. RESULTS: Fifty-two (96%) of the 54 nodules were successfully segmented, allowing their volume to be calculated. Repeatability was high: There was no variation in the nine measurements of 35 (67%) of the 52 nodules. The coefficient of variation for the remaining 17 nodules (33%) was 2.26%. Bland and Altman 95% limits of acceptability, calculated on the basis of log-transformed data, yielded a maximum software measurement error of 6.38% of the previous volume measurement. Doubling time ranged from 4 to 188 years for the 13 benign nodules and from 37 to 216 days for the nine malignant nodules. CONCLUSION: Software volumetric analysis yielded repeatable estimates for 96% of the nodules examined. All software-calculated doubling times were in keeping with the benign or malignant nature of the nodules.
Keywords:
本文献已被 PubMed 等数据库收录!
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号