动态轮廓眼压计与Goldmann压平眼压计及非接触眼压计测量眼压的对比研究
引用本文:程玲艳,崔娟莲,段宣初.动态轮廓眼压计与Goldmann压平眼压计及非接触眼压计测量眼压的对比研究[J].眼科,2011,20(1):33-37.
作者姓名:程玲艳  崔娟莲  段宣初
作者单位:1. 江苏省无锡市第二人民医院眼科
2. 中南大学湘雅二医院眼科,长沙,410011
摘    要:目的探讨动态轮廓眼压计(DCT)与Goldmann压平眼压计(GAT)及非接触眼压计(NCT)测量眼压的准确性,并比较三种眼压计测量结果与中央角膜厚度(CCT)的相关性。设计前瞻性、比较性病例系列。研究对象连续选取90例(90眼)10~76岁正常人。方法采用KONAN非接触式角膜内皮镜测量CCT后,对所有入选者单眼以随机顺序采用Pascal型DCT、GAT及Topcon型NCT测量眼压。测量结果两两比较,并将眼压值与CCT进行直线回归分析。主要指标眼压值,Pearson相关系数。结果 90例正常人DCT眼压平均值(17.33±2.71 mm Hg)明显高于GAT(14.27±2.81 mm Hg)(P=0.000)及NCT(14.67±2.93 mm Hg)(P=0.000),平均差异分别为(3.06±2.01)mm Hg和(2.67±2.20)mm Hg;GAT与NCT之间平均差异为(-0.39±2.29)mm Hg(P=0.105)。DCT与GAT眼压值之间相关系数r=0.736(P=0.000);与NCT眼压值之间相关系数r=0.699(P=0.000)。GAT、NCT眼压值与CCT均明显相关(r=0.370,P=0.000;r=0.508,P=0.000);DCT眼压值与CCT无明显相关性(r=0.051,P=0.639)。DCT和GAT的差值与年龄无明显相关性(r=0.064,P=0.052)。结论 DCT测量的眼压值虽高于GAT及NCT,但不受CCT的影响,可能较GAT和NCT测量的眼压值更接近真实值。

关 键 词:眼压  眼压计  中央角膜厚度

Comparative study of intraocular pressure measurement by dynamic contour tonometry, Goldmann applanation tonometry and non-contact tonometry in normal subjects
CHENG Ling-yan,GUI Juan-lian,DUAN Xuan-chu.Comparative study of intraocular pressure measurement by dynamic contour tonometry, Goldmann applanation tonometry and non-contact tonometry in normal subjects[J].Ophthalmology in China,2011,20(1):33-37.
Authors:CHENG Ling-yan  GUI Juan-lian  DUAN Xuan-chu
Institution:CHENG Ling-yan,CUI Juan-lian,DUAN Xuan-chu.The Second Xiangya Hospital,Central South University,Changsha 410011,China
Abstract:Objective To compare the accuracy of introcular pressure(IOP) measured by Pascal dynamic contour tonometry(DCT),Goldmann applanation tonometry(GAT) and Topcon non-contact tonometry(NCT),and to analyze the relationship between central corneal thickness(CCT) and IOP value measured by DCT,GAT and NCT.Design Prospective,comparative case series.Participant 90 healthy individuals aged 10 to 76 years old.Method Chose continuously 90 single eyes of 90 healthy individuals for IOP measuring.After CCT measurement usin...
Keywords:intracular pressure  tonometry  central corneal thickness  