Oxygenator exhaust capnography as an index of arterial carbon dioxide tension during cardiopulmonary bypass using a membrane oxygenator
Authors:O'Leary M J  MacDonnell S P  Ferguson C N
Institution:Department of Anaesthesia, St Bartholomew's Hospital, West Smithfield, London EC1A 7BE, UK
Abstract:We have studied the relationship between the partial pressure of carbon dioxide in oxygenator exhaust gas (PECO2) and arterial carbon dioxide tension (PaCO2) during hypothermic cardiopulmonary bypass with non- pulsatile flow and a membrane oxygenator. A total of 172 paired measurements were made in 32 patients, 5 min after starting cardiopulmonary bypass and then at 15-min intervals. Additional measurements were made at 34 degrees C during rewarming. The degree of agreement between paired measurements (PaCO2 and PECO2) at each time was calculated. Mean difference (d) was 0.9 kPa (SD 0.99 kPa). Results were analysed further during stable hypothermia (n = 30, d = 1.88, SD = 0.69), rewarming at 34 degrees C (n = 22, d = 0, SD = 0.84), rewarming at normothermia (n = 48, d = 0.15, SD = 0.69) and with (n = 78, d = 0.62, SD = 0.99) or without (n = 91, d = 1.07, SD = 0.9) carbon dioxide being added to the oxygenator gas. The difference between the two measurements varied in relation to nasopharyngeal temperature if PaCO2 was not corrected for temperature (r2 = 0.343, P = < 0.001). However, if PaCO2 was corrected for temperature, the difference between PaCO2 and PECO2 was not related to temperature, and there was no relationship with either pump blood flow or oxygenator gas flow. We found that measurement of carbon dioxide partial pressure in exhaust gases from a membrane oxygenator during cardiopulmonary bypass was not a useful method for estimating PaCO2.
Keywords:
本文献已被 PubMed Oxford 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号