拜安捷TM血糖监测仪测定末梢毛细血管血糖与全自动生化分析仪测定静脉血浆血糖的临床对比研究
引用本文:姚军,高妍,纪立农,郭晓蕙,陈澜,王波澜. 拜安捷TM血糖监测仪测定末梢毛细血管血糖与全自动生化分析仪测定静脉血浆血糖的临床对比研究[J]. 中华内分泌代谢杂志, 2006, 22(3): i0009-i0012
作者姓名:姚军  高妍  纪立农  郭晓蕙  陈澜  王波澜
作者单位:1. 100034,北京大学第一医院内分泌科
2. 北京大学人民医院内分泌科
摘    要:目的 评价拜安捷^TM血糖监测仪的精确性及其测定毛细血管血糖(capillary blood glucose,CBG)与全自动生化分析仪测定静脉血浆血糖(venous plasma glucose,VPG)的相关性。方法 188例糖尿病或非糖尿病受试者在空腹及标准餐后30、60、120min测定VPG同时用拜安捷^TM血糖监测仪测定CBG。用10台拜安捷^TM血糖监测仪分别用三个批号的试纸对低、中、高浓度水平的核对液进行测定计算其批内差异和批间差异。结果 在空腹及进标准餐后不同时间点,CBG与VPG均有良好的相关性,r均在0.950以上。误差分析图显示98.94%的数据落在A区。在不同血糖浓度下,CBG与VPG的相对差值均在5%以下。在低浓度、正常浓度及高浓度范围内,采用不同批号试纸检测,批内差异和批间差异CV值均〈5%。结论 拜安捷^TM血糖监测仪的准确性和精确性高,在空腹和进餐后不同时间及不同血糖浓度下均能较准确地反映血糖的真实水平。

关 键 词:糖尿病 血糖仪 毛细血管血糖 静脉血浆血糖

Comparison of periphery capillary whole blood glucose using BREEZE TM blood glucose meter and venous plasma glucose using laboratory autoanalyzer
YAO Jun,GAO Yan,JI Li-nong,GUO Xiao-hui,CHEN Lan,WANG Po-lan. Comparison of periphery capillary whole blood glucose using BREEZE TM blood glucose meter and venous plasma glucose using laboratory autoanalyzer[J]. Chinese Journal of Endocrinology and Metabolism, 2006, 22(3): i0009-i0012
Authors:YAO Jun  GAO Yan  JI Li-nong  GUO Xiao-hui  CHEN Lan  WANG Po-lan
Affiliation:Department of Endocrinology, Peking University First Hospital, Beijing 100034
Abstract:Objective To evaluate the precision of BREEZE~(TM) glucose monitor and correlation of glucose measurements between fingertip capillary whole blood glucose(CBG)using BREEZE~(TM) glucose monitor and venous plasma glucose(VPG)using autoanalyzer.Methods All samples of venous plasma and fingertip blood from 188 diabetes or non-diabetes patients were detected for glucose level at fasting,30,60 and 120 min postprandial.BREEZE~(TM) glucose monitor and autoanalyzer measured CBG and VPG respectively.Intra-and inter- coefficients of variations were determined using 10 BREEZE~(TM) glucose monitors with normal,slight high and high glucose levels for the three lots of strips.Results The correlation coefficients between CBG and VPG were all higher than 0.950 at fasting and different postprandial time.98.94% of all measurements were in the A zone when using error-grid analysis.The relevant differences between CBG and VPG were less than 5% at different blood glucose concentrations.The intra-and inter-coefficients of variations of blood glucose values at different blood glucose concentrations using different lots of strips were within 5%.Conclusion BREEZE~(TM) glucose monitor provides high accurate and precise glucose readings on fasting and different postprandial time points over a variety of blood glucose concentrations.
Keywords:Diabetes mellitus   Glucose meters   Capillary whole blood glucose   Venous plasma glucose