An Implanted Closed-loop Chip System for Heart Rate Control: System Design and Effects in Conscious Rats
Authors:Yuxuan Zhou  Yuan Yuan  Juan Gao  Ling Yang  Feng Zhang  Guoqing Zhu  Xingya Gao
Affiliation:aDepartment of Biomedical Engineering, Nanjing Medical University, Nanjing, 210029, China;bDepartment of Physiology, Nanjing Medical University, Nanjing, 210029, China
Abstract:

Objective

To evaluate the efficiency of an implanted chip system for the control of heart rate (HR).

Methods

The HR was recorded in six conscious Sprague-Dawley (SD) rats. An implanted chip system was designed to regulate the HR by stimulating the right cervical vagus nerve according to the feedback of real time HR. Each rat was subjected to 30-min regulation and 30-min recovery. The change of HR during the regulation period was compared with the control. The ECG was recorded during the experiment for 24 h.

Results

The ECG signals were successfully recorded during the experiment. The HR was significantly decreased during the period of regulation compared with control (-79.3 ±34.5, P < 0.01, n = 6) and then recovered to normal after regulation.

Conclusion

The described implanted chip system can regulate the HR to a designated set point.
Keywords:closed-loop regulation   implanted chip system   set point   heart rate   vagus nerve stimulation   Remote-controlled animal