首页 | 本学科首页   官方微博 | 高级检索  
     


Inadvertent positioning of pacemaker leads in the pericardium
Authors:Berenji Kambeez  Nerheim Pamela  Olshansky Brian
Affiliation:Department of Cardiology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
Abstract:A patient had a dual chamber pacemaker with endocardial leads implanted chronically. The lead position on chest X ray and the ECG pattern indicated lead malposition, but a CT scan and transesophageal echocardiography were nondiagnostic. Venography indicated that both leads were in the mediastinal and pericardial space.
Keywords:pacemakers    lead malposition    pericardium    venogram
本文献已被 PubMed 等数据库收录!