首页 | 本学科首页   官方微博 | 高级检索  
     

咳停合剂微生物限度检查法的建立与验证
引用本文:杨征,陈雅,徐果. 咳停合剂微生物限度检查法的建立与验证[J]. 中国药业, 2014, 0(17): 34-35
作者姓名:杨征  陈雅  徐果
作者单位:中国人民解放军第三军医大学野战外科研究所大坪医院药剂科,重庆400042
摘    要:目的:建立并验证咳停合剂的微生物限度检查方法。方法采用常规法、薄膜过滤法检查咳停合剂的微生物限度,并依据2010年版《中国药典(二部)》收载的方法进行验证。结果咳停合剂供试品试验菌株的回收率均高于70%,证明无抑菌现象。结论采用常规法与薄膜过滤法检查咳停合剂微生物限度均可行。

关 键 词:咳停合剂  微生物限度检查  常规法  薄膜过滤法  验证

Establishment and Validation of Microbial Limit Test of Keting Mixture
Yang Zheng,ChenYa,Xu Guo. Establishment and Validation of Microbial Limit Test of Keting Mixture[J]. China Pharmaceuticals, 2014, 0(17): 34-35
Authors:Yang Zheng  ChenYa  Xu Guo
Affiliation:(Department of Pharmacy, Research Institute of Field Surgery, Daping Hospital, Third Military Medial University, Chongqing, China 400042)
Abstract:Objective To establish and validate a microbial limit test for Keting Mixture. Methods The conventional method and the membrane-filter procedure were adopted to detect the microbial limit of Keting Mixture and the validation was performed according to the method in the Chinese Pharmacopoeia ( version 2010, Part 2 ) . Results The recovery rate of experimental bacterial strain in Keting Mixture sample exceeded 70%, which proved that no bacteriostasis phenomena existed. Conclusion Adopting the conventional method and the membrane-filter procedure is feasible for detecting the microbial limit of Keting Mixture.
Keywords:Keting Mixture  microbial limit test  conventional method  membrane-filter procedure  validation
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号