首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of specific protein assays in biological fluids by radial immunodiffusion and laser nephelometer.
Authors:G Shulman
Affiliation:Department of Clinical Pathology, The Milton S. Hershey Medical Center of The Pennsylvania State University, Hershey, Pennsylvania 17033, USA
Abstract:Laser immuno-nephelometry is presently being used for measuring concentrations of specific proteins in biological fluids. Our findings have not substantiated the highly correlative results with radial immunodiffusion that have been reported. The reason for our poor correlation was found in the lack of uniformity in expressing concentrations of the different specific proteins in the calibrating standards. One value of the laser nephelometer was the greater precision obtained when comparing results to those measured by radial immuno-diffusion. Also, the laser was found to be more sensitive for measuring low level concentrations.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号