血气分析仪与全自动生化仪相同检测项目可比性探讨
引用本文:杨惠聪,林洁,方春香,王美玉.血气分析仪与全自动生化仪相同检测项目可比性探讨[J].检验医学与临床,2013(16):2072-2073.
作者姓名:杨惠聪  林洁  方春香  王美玉
作者单位:1. 福建医科大学附属漳州市医院 检验科,福建漳州,363000
2. 福建医科大学附属漳州市医院 胸心外科,福建漳州,363000
摘    要:目的探讨血气分析仪与生化分析仪检测动脉血与静脉血样本中血钾(K+)、血钠(Na+)、血氯(Cl-)以及血总二氧化碳(TCO2)的相关性。方法分别用ABL800血气分析仪与OLYMPUS AU5400全自动生化分析仪对36例患者的动脉全血和血清样本进行K+、Na+、Cl-、TCO2的检测,对检测结果进行相关分析和配对t检验。结果两种仪器测定K+、Na+、Cl-、TCO2的结果均有显著相关性(r=0.717,r=0.832,r=0.721,r=0.919,P<0.01)。两种仪器测定TCO2的结果差异无统计学意义(P>0.05),ABL800血气分析仪测定K+、Na+、Cl-的结果显著高于OLYMPUS AU5400全自动生化分析仪(P<0.05)。结论 ABL800血气分析仪与AU5400生化分析仪K+、Na+、Cl-、TCO2检测结果相关性好,但仍存在一定的差异。

关 键 词:血气分析仪  生化分析仪  血钾  血钠  总二氧化碳

Analysis of the same test items between blood gas analyzer and biochemical analyzer
YANG Huicong , LIN Jie , FA NG Chunxiang , WA NG Meiyun.Analysis of the same test items between blood gas analyzer and biochemical analyzer[J].Laboratory Medicine and Clinic,2013(16):2072-2073.
Authors:YANG Huicong  LIN Jie  FA NG Chunxiang  WA NG Meiyun
Institution:1.Department of Clinical Laboratory;2.Department of Cardiovascular Surgical,Affiliated Hospital of Fujian Medical University,Zhangzhou,Fujian 363000,China)
Abstract:Objective To explore the correlation and possible differences of blood potassium,sodium,chlorine and total carbon dioxide measured by blood gas analyzer and biochemical analyzer.Methods The levels of artery or vein blood potassium,sodium,chlorine and total carbon dioxide from 36 patients were examined by ABL800 blood gas analyzer and OLYMPUS AU5400 automated biochemical analyzer,and Pearson correlation analysis and paired t-test were employed to analyze the results.Results Significant correlation was found between results of blood potassium,sodium,chlorine and total carbon dioxide of blood gas analyzer and biochemical analyzer(r=0.717,r=0.832,r=0.721,r=0.919,P0.01).There was no significant difference in total carbon dioxide concentration detected by these two instruments(P0.05),while potassium,sodium and chlorine concentration detected by ABL800 blood gas analyzer was significantly higher than those detected by OLYMPUS AU5400 biochemical analyzer(P0.05).Conclusion The results of the two analyzers could have significant correlation,but also have some difference.
Keywords:blood gas analyzer  biochemical analyzer  blood potassium  blood sodium  blood total car-bon dioxide