首页 | 本学科首页   官方微博 | 高级检索  
     


Chronic evaluation of the Cleveland Clinic CorAide left ventricular assist system in calves
Authors:Fukamachi Kiyotaka  Ochiai Yoshie  Doi Kazuyoshi  Massiello Alex L  Medvedev Alexander L  Horvath David J  Gerhart Renee L  Chen Ji-Feng  Krogulecki Alexandra Y  Takagaki Masami  Howard Michael W  Kopcak Michael W  Golding Leonard A R
Affiliation:Department of Biomedical Engineering and Biological Resources Unit, Lerner Research Institute, Cleveland Clinic Foundation, Cleveland, Ohio, U.S.A.
Abstract:The Cleveland Clinic CorAide left ventricular assist system is based on a third-generation, implantable, centrifugal pump in which a rotating assembly is suspended fully. To evaluate chronic in vivo system performance and biocompatibility, the CorAide blood pump was implanted in 18 calves for either 1 month or 3 months. Hemodynamics were stable in all calves with a mean pump flow of 5.9 +/- 1.2 L/min and a mean systemic arterial pressure of 98 +/- 5 mm Hg. There were no incidences of bleeding, organ dysfunction, or mechanical failure in any of the 18 calves. Hemolysis occurred in only 1 calf due to outflow graft stenosis. Thrombus inside the pump, seen in 4 of the first 6 cases, was totally eliminated by a final redesign in the remaining cases, including the last 6 implants conducted without anticoagulation therapy. The CorAide blood pump demonstrated good biocompatibility and reliable, effective system performance.
Keywords:Centrifugal pump    Left ventricular assist device    Anticoagulation    Hemolysis