Assessment of a needle-tracking system for bipolar radiofrequency ablation
Authors:Masashi Hirooka  Yohei Koizumi  Yusuke Imai  Hironori Ochi  Yoshiko Nakamura  Osamu Yoshida  Atsushi Hiraoka  Yoshio Tokumoto  Masanori Abe  Yoichi Hiasa
Institution:1.Department of Gastroenterology and Metabology,Ehime University Graduate School of Medicine,Toon,Japan
Abstract:

Purpose

The purpose of this study was to assess the accuracy and efficacy of a needle-tracking system in phantom and clinical studies using bipolar electrodes.

Methods

To observe the tip of the electrode, a needle-tracking system with a volume navigation system was used. In the phantom study, the electrode was inserted at various angles and the error was verified. In the clinical study, 21 nodules close to extrahepatic organs or major vessels were enrolled between May and October 2014. After puncturing with the needle-tracking system, computed tomography (CT) was performed. The distances between the electrode tip and extrahepatic organs or major vessels were measured on both B-mode ultrasound (US) and CT. By comparing these distances, the accuracy of this system was evaluated.

Results

In the phantom study, the deviation between the tip of the electrode and the virtual tip of the electrode was analyzed. The median values were within 2 mm at each puncture angle. In the clinical study, the difference between B-mode US and CT was less (mean value 1.17 ± 1.76 mm; range 0–3.5 mm).

Conclusion

The needle-tracking system is an accurate and useful system for bipolar radiofrequency ablation.