首页 | 本学科首页   官方微博 | 高级检索  
     


PRELOAD AND TORQUE REMOVAL EVALUATION OF THREE DIFFERENT ABUTMENT SCREWS FOR SINGLE STANDING IMPLANT RESTORATIONS
Authors:Rafael Augusto Stüker  Eduardo Rolim Teixeira  Jo?o Carlos Pinheiro Beck  Nilza Pereira da Costa
Abstract:Several authors still consider the mechanical problems of fracture and component loosening as the main causes of failure of implant-supported restorations. The purpose of this in vitro study was to compare the preload of three types of screw for transmucosal abutment attachment used in single implant-supported prosthesis through strain gauge and removal torque measurements. Three external hex fixtures were used, and each received a transmucosal abutment (Cera One®), which was fixed to the implant with its respective screw: Group A- gold screw, Group B- titanium screw and Group C- surface-treated titanium screw (Ti-Tite®). Ten screws of each type were attached applying a 30.07±0.28 Ncm torque force and maintained in position for 5 minutes. After this, the preload values were measured using strain gauges and a measurement cell. Gold screws presented higher preload values (131.72±8.98 N), followed by surface-treated titanium screws (97.78±4.68 N) and titanium screws (37.03±5.69 N). ANOVA (p<0.05) and Tukey''s test (p<0.05) were applied. Statistically significant differences were found among the groups for both preload and removal torque values. In conclusion, gold screws may be indicated to achieve superior longevity of the abutment-implant connection and, consequently, prosthetic restoration due to greater preload values yielded.
Keywords:Dental implants   Dental prosthesis   Torque   Screw   Abutment
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号