The cooling effect of liquid flow on the focussed ultrasound-induced heating in a simulated foetal brain
Authors:Vella Gilbert J  Humphrey Victor F  Duck Francis A  Barnett Stanley B
Affiliation:School of Biomedical Sciences, The University of Sydney, Sydney, Australia. G.Vella@fhs.usyd.edu.au
Abstract:There is a need to investigate the thermal effects of diagnostic ultrasound (US) to assist the development of appropriate safety guidelines for obstetric use. The cooling effect of a single liquid flow channel was measured in a model of human foetal brain and skull bone heated by a focussed beam of simulated pulsed spectral Doppler US. Insonation conditions were 5.7 micros pulses, repeated at 8 kHz from a focussed transducer operating with a centre frequency of 3.5 MHz, producing a beam of -6 dB diameter of 3.1 mm at the focus and power outputs of up to 255 +/- 5 mW. Brain perfusion was simulated by allowing distilled water to flow at various rates in a 2 mm diameter wall-less channel in the brain soft tissue phantom material. This study established that the cooling effect of the flowing water; 1. was independent of the acoustic source power, 2. was more effective close to the flow channel, for example, there was a marked cooling at a distance of 1 mm and negligible cooling at a distance of 3 mm from the channel; and 3. initially increased at low flow rates, but further increase above normal perfusion had very little effect.