首页 | 本学科首页   官方微博 | 高级检索  
     

计算机导航系统和C臂机透视引导颈椎椎弓根螺钉内固定技术的临床对比研究
引用本文:Tian W,Liu YJ,Liu B,Li Q,Hu L,Li ZY,Yuan Q. 计算机导航系统和C臂机透视引导颈椎椎弓根螺钉内固定技术的临床对比研究[J]. 中华外科杂志, 2006, 44(20): 1399-1402
作者姓名:Tian W  Liu YJ  Liu B  Li Q  Hu L  Li ZY  Yuan Q
作者单位:100035,北京积水潭医院脊柱外科
摘    要:目的探讨颈椎(C1-C7)椎弓根螺钉内固定的可行性和两种不同置钉方法的精确性。方法通过术后进行X线摄片、CT或术后Iso-C断层扫描等方法,判断C臂机透视引导下置入的145枚和三维导航系统(CT三维导航或Iso—C术中三维导航)辅助下187枚颈椎椎弓根螺钉病例的置钉准确性。结果三维导航系统引导组螺钉置入满意率(97.9%,183/187)明显高于C臂机透视引导组(91.7%,133/145)(χ2=6.705,P=0.010)。且三维导航系统引导组位置不满意的4枚螺钉均发生于早期病例,导航系统使用熟练后未再出现置钉不满意的病例。两组病例均未出现明显的神经血管损伤并发症(C臂机透视引导组2例术后一过性上肢疼痛)。对三维导航系统引导组中25例进行术中导航操作时间和导航精确性的监测。CT三维导航术中工具注册和参考点照合时间平均3.5min(2—8rain),位置误差率平均0.31mm(0.12—0.56mm,导航仪自动计算);Iso—C术中三维导航图像采集和传输时间平均6.2min(5—7min),每颗椎弓根螺钉定位针置入所需时间平均2min(1—3.5min);术中只需进行2次C臂机透视印证螺钉定位针和螺钉置入的准确性。结论采用三维导航系统辅助,能显著提高椎弓根螺钉置入的精确性,其操作技术有待完善。

关 键 词:颈椎 内固定器 治疗 计算机辅助
收稿时间:2006-03-02
修稿时间:2006-03-02

Clinical contrast of cervical pedicle screw fixation assisted by C-arm fluoroscopy or 3D navigation system
Tian Wei,Liu Ya-jun,Liu Bo,Li Qin,Hu Lin,Li Zhi-yu,Yuan Qiang. Clinical contrast of cervical pedicle screw fixation assisted by C-arm fluoroscopy or 3D navigation system[J]. Chinese Journal of Surgery, 2006, 44(20): 1399-1402
Authors:Tian Wei  Liu Ya-jun  Liu Bo  Li Qin  Hu Lin  Li Zhi-yu  Yuan Qiang
Affiliation:Department of Spine Surgery, Beijing Jishuitan Hospital, Beijing 100035, China
Abstract:Objective To evaluate the feasibility and accuracy of cervical(C_2-C_7)pedicle screw fixation assisted by C-arm fluoroscopy or 3D navigation system(CT-based navigation system or Iso-C 3D navigation system).Methods One hundred and forty-five cervical pedicle screws inserted with C-arm fluoroscopy and 187 pedicle screws inserted with 3D navigation system were observed by postoperative CT or Iso-C 3D scan.The process of navigation was investigated.Results In the 145 screws inserted with C-arm fluoroscopy,133 screws are accurate(91.7%),and in the 187 screws inserted by 3D navigation system, 183 screws are accurate(97.9%)(x~2=6.705,P=0.010),and the 4 misplaced screws occurred inthe early stage of navigation system application.The process of navigation was investigated in twenty-five patients.With CT-based navigation,the mean time for registration and surface matching was 3.5 minutes; with Iso-C 3D navigation,the mean time for images collection and transfer was 6.2 minutes.The mean time for screw-marker insertion was 2 minutes.Conclusions 3D navigation system can increase accuracy of cervical pedicle screw fixation obviously.
Keywords:Cervical vertebrae   Internal fixators    Therapy, computer-assisted
本文献已被 CNKI 维普 万方数据 PubMed 等数据库收录!