Microdislodgment: |
| |
Authors: | KAN KIKUCHI HARUHIKO ABE TOSHIHISA NAGATOMO YASUHIDE NAKASHIMA |
| |
Institution: | From the Second Department of Internal Medicine, University of Occupational and Environmental Health, Kitakyushu, Japan |
| |
Abstract: | High impedance tined steroid-eluting leads, Medtronic CapSure Z family, incorporating a small surface (1.2 mm2) electrode made of porous platinum material are designed to reduce battery current drain. However, in some prior studies, an increased incidence of microdislodgment with this lead was reported thought to be due to the reduced electrode surface area. This article reports the experience that ventricular pacing failure due to microdislodgment occurred after CapSure Z lead implantation and the previous literature is reviewed. (PACE 2003; 26Pt. I]:1541–1543) |
| |
Keywords: | high impedance small surface pacing failure microdislodgment |
|
|