首页 | 本学科首页   官方微博 | 高级检索  
检索        


Microdislodgment:
Authors:KAN KIKUCHI  HARUHIKO ABE  TOSHIHISA NAGATOMO  YASUHIDE NAKASHIMA
Institution:From the Second Department of Internal Medicine, University of Occupational and Environmental Health, Kitakyushu, Japan
Abstract:High impedance tined steroid-eluting leads, Medtronic CapSure Z family, incorporating a small surface (1.2 mm2) electrode made of porous platinum material are designed to reduce battery current drain. However, in some prior studies, an increased incidence of microdislodgment with this lead was reported thought to be due to the reduced electrode surface area. This article reports the experience that ventricular pacing failure due to microdislodgment occurred after CapSure Z lead implantation and the previous literature is reviewed. (PACE 2003; 26Pt. I]:1541–1543)
Keywords:high impedance                        small surface                        pacing failure                        microdislodgment
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号