Airway and body surface sensors for triggering in neonatal ventilation |
| |
Authors: | J John,LJ Bjö rklund,NW Svenningsen,B Jonson |
| |
Affiliation: | Department of Clinical Physiology and Neonatal Intensive Care Unit of the Department of Paediatrics;Department of Paediatrics, University Hospital, Lund, Sweden |
| |
Abstract: | Failure of neonatal patient triggered ventilation may reflect a delay in delivery of flow relative to the inspiratory effort of the infant. Transmission of diaphragmatic contraction to the sensor site (patient delay) and further transmission to and within the sensing device (device delay) both contribute to the delay in triggering. Patient and device delays were studied for different sensing systems in 36 infants, 24 of whom were intubated. Device delay was long (<40 ms) with a conventional apnoea monitor compared with sensors placed at the airway opening (2 ms), the inspiratory (12 ms) and expiratory (3 ms) pressure transducers of the ventilator, the Graseby capsule (8 ms), strain gauges (3 ms) and oesophageal pressure (6 ms). In near normal infants, the sum of patient and device delays for the latter sensors was less than 20 ms and a minor component of the total delay. However, in severe lung disease the total delay may be more than 100ms even for airway sensors. |
| |
Keywords: | Infant mechanical ventilation newborn respiratory distress syndrome |
|
|