首页 | 本学科首页   官方微博 | 高级检索  
     


Aortic Valve Replacement with Starr-Edwards Valves over 14 Years
Authors:W.H. Wain  P.J. Drury  D.N. Ross
Affiliation:From the Department of Surgery, Cardiothoracic Institute, London, England
Abstract:Three hundred thirteen patients underwent aortic valve replacement with 319 Starr-Edwards caged-ball prostheses and have been followed for 1 to 14 years. Hospital mortality (24.8%) and first-year mortality (4.8%) suggest that this is a high-risk group. Overall postoperative patient survival was 37.1%, with 18.8% free from any event, at 14 years. Thromboembolism was the most significant single event (a probability of 18.7% at 14 years), and the Series 2300/2320 valves were associated with a probability of hemolysis of nearly 80% over 13 years.Patients who underwent valve replacement before 1973 had a significantly greater probability of late death (31.1%) and of complications (47.1%) during the first 5 years. Those patients undergoing replacement after 1973 had a significantly greater probability of thromboembolic episodes (15.4%).Starr-Edwards caged-ball prostheses provide an acceptable valve replacement for high-risk patients. However, the overall complication rate of 81.2% over 14 years, compared with 78% for homograft valves for the same period, does not support the adoption of this prosthesis as the valve of first choice at this hospital.
Keywords:Address reprint requests to Mr. Ross   Department of Surgery   Cardiothoracic Institute   2 Beaumont St   London W1N 2DX   England
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏