Use of magnetisation transfer contrast to improve cerebral 3D MR angiography |
| |
Authors: | V. Dousset J. M. Franconi P. Degrèze J. Balderrama F. Lexa J. M. Caillé |
| |
Affiliation: | (1) Service de Neuroradiologie, Hopital Universitaire Pellegrin, Place Amélie Raba-Léon, F-33076 Bordeaux Cedex, France;(2) Siemens, Paris, France |
| |
Abstract: | To improve 3D-time of flight (3D-TOF) magnetic resonance angiography, we used magnetisation transfer (MT) to increase the contrast between flowing blood and stationary tissues. With a 1.5 KHz off-resonance radio-frequency MT applied for 16 ms at a maximum power within specific absorption rate (SAR) limits, a 37% decrease in the signal of white matter was obtained, whereas the signal from flowing blood decreased by only 8%. An improvement in maximum intensity projection (MIP) image quality was obtained all MT-3DTOF studies on seven volunteers using progressively more powerful MT pulses. Routine clinical use of MT-3DTOF appears promising and can be achieved at any strength field. |
| |
Keywords: | Magnetic resonance imaging techniques Magnetic resonance angiography Magnetisation transfer contrast |