Abstract: | Water content‐based electrical properties tomography (wEPT) can retrieve electrical properties (EPs) from water content maps, thereby eliminating the need for B1 field measurement in the traditional magnetic resonance electrical properties tomography method. The wEPT is performed by conventional MR scanning, such as T1 ‐weighted spin‐echo imaging, and thus can be directly applied to clinical settings. However, the random noise propagation involved in wEPT causes inaccuracy in EP mapping. To guarantee the EP estimates desired for clinical practice, this study statically investigates the noise‐specific uncertainty of wEPT through probability density function models. We calculated the probability distribution of EP maps with different noise levels and examined the effects of scan parameters on reconstruction accuracy with various flip angles (FAs) and repetition time (TR) settings. The theoretical derivation was validated by Monte Carlo simulations and human imaging experiment at 3 T. Results showed that a serious deviation could occur in tissues with large conductivity value at a low signal‐to‐noise ratio and quantitatively demonstrate that such deviation could be mitigated by increased FAs or TRs. This study provided useful information for the setup of scan parameters, evaluation of accuracy of the wEPT under specific SNR levels, and promote its clinical applications. |