A study of calcium ion-selective PVC membrane electrode based on neutral carrier N,n,n’,n’-tetracyclo-3-oxapetanediamide |
| |
Authors: | Ma Yi-long Rao Xiao-hong Zhong Shi-ming Ren Shu Yu Tian-xiang Zhen Qiong |
| |
Affiliation: | (1) Institute of Environmental Medicine, Tongji Medical University, Wuhan;(2) Department of Chemistry, University of Wuhan, Wuhan |
| |
Abstract: | Summary A calcium ion-selective PVC membrane electrode based on neutral carrier n,n,n’,n’-tetracyclohexyl-3-oxapetanediamide, using di-(2-ethylhexyl) phthalate as the plasticizer and potassium tetrakis (4-chlorophenyl) borate as the additive is reported in this paper. The ion selective membrane consists of 1 wt % of the Ca2+ selective ligand, 65 wt % of the plasticizer, 1 wt % of the additive and 33 wt % of poly (vinyl chloride) powder. The electrode has the linear response range of 2.0 × 10™7 —10™1 mol/L with the Nernstian slope of 28 mV/decade at 25 °C and the detection limit of 2.0 × 10™8 mol/L. The response time of the calcium ion-selective electrode is 4s as the concentration of calcium ion is rapidly shifted from 10™5 to 10™4 mol/L. The potential stability and reproducibility are good. The free calcium in blood serum was determined by the calcium ion-selective electrode with satisfactory results. |
| |
Keywords: | calcium ion-selective electrode neutral carrier PVC membrane calcium determination in serum |