首页 | 本学科首页   官方微博 | 高级检索  
     


Noninvasive Assessment of Ventriculo-Atrial Conduction and Early Experience with the Tachycardia Termination Algorithm in Pacemaker-Mediated Tachycardia
Authors:JOHN M. FONTAINE  JAMES D. MALONEY  LON W. CASTLE  VICTOR A. MORANT
Affiliation:Cleveland Clinic Foundation, Cleveland, Ohio
Abstract:Since the advent of physiologic dual chamber pacing systems, pacemaker-mediated tachycardia (PMT) has occurred and the need for invasive measurement of ventriculo-atrial conduction (VAC) has arisen. The variability in VAC and the potential for PMT often make it necessary to assess for the presence or absence of VAC at different points in time. We noninvasively evaluated 20 pacemaker patients for the presence or absence of VAC. We compared ventriculo-atrial conduction time (VACT) obtained with the atrial sense event maker with that obtained from Holter monitoring and invasive methods. The incidence of spontaneous (S) and induced (I) PMT and the efficacy of the tachycardia termination algorithm (TTA) was assessed. Fourteen of 20 had VAC with invasive or noninvasive methods. Twelve of 19 had PMT (63%); three were sustained (greater than 15 beats). We conclude that VACT assessed with the atrial sense event marker (ASEM) yielded a high correlation when compared to the Holter monitor data obtained utilizing our methodology. PMT is commonly a nonsustained (less than 15 beats) event, and the TTA is effective in sustained PMT. Myopotential sensing, atrial premature contractions and loss of atrial capture are common mechanisms in the initiation of PMT.
Keywords:dual chamber pacemaker    physiologic pacing    pacemaker-mediated tachycardia    ventriculo-atrial conduction    tachycardia termination algorithm
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号