首页 | 本学科首页   官方微博 | 高级检索  
     


A Preliminary Flow Visualization Study in a Multiple Disk Centrifugal Artificial Ventricle
Authors:G.E. Miller  M. Madigan  R. Fink
Affiliation:Bioengineering Program, Texas A&M University, College Station, Texas, U.S.A.
Abstract:Abstract: Flow patterns in a multiple disk centrifugal pump were analyzed so that the device could be incorporated as a ventricular assist or a bridge-to-transplant device. The pump operates either in pulsatile or steady flow modes with the ability to change modes within a fraction of a second. The pump was tested on a mock circulatory system consisting of an arterial fluid capacitor, a systemic resistor, and a venous capacitor. Arterial volume flow rate, arterial pressure, inlet (venous) pressure, and pump rotation speed are continually monitored. A glycerin/ water solution is used as a blood analog. Flow visualization was performed with a 3 mW yellow laser, sheet lens, neutrally buoyant amberlite particles, and both still and motion picture photography. Flow patterns matched theoretical predictions very well; inlet flow spread radially outward through the disk annular spaces while propelled by shear and centrifugal forces.
Keywords:Disk pump    Ventricular assist device    Centrifugal pump
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号