大、小直径钛网对终板-钛网界面应力分布影响的比较
引用本文:苏再发,贾连顺,陈雄生,李家顺,张美超,李鉴软,赵卫东,吴培增. 大、小直径钛网对终板-钛网界面应力分布影响的比较[J]. 颈腰痛杂志, 2011, 32(5): 323-327
作者姓名:苏再发  贾连顺  陈雄生  李家顺  张美超  李鉴软  赵卫东  吴培增
作者单位:1. 福建省泉州市中医院骨科,泉州,362000
2. 上海第二军医大学附属长征医院骨科,上海,200003
3. 南方医科大学解剖教研室,广东省医学生物力学重点实验室,广州,510515
基金项目:泉州市科技优秀人才专项经费资助项目,泉州市局资助项目
摘    要:目的 比较大、小直径钛网植骨对颈椎椎体次全切除减压终板-钛网界面应力分布的影响.方法 利用Ansys 9.0软件的建模功能,建立大、小直径钛网植骨钢板固定手术模型.分别对模型施加80 N预载荷及1.8 Nm力矩,使其产生轴向压缩、前屈和后伸运动,选取终板-钛网界面7个接触点,提取各个点的von Mise.应力,然后进行...

关 键 词:颈椎  椎体次全切除  终板  钛网  下沉  生物力学  有限元法

A finite element analysis of comparision study on mechanical changes on cage-endplate interface between different size of the titanium mesh cages with larger or small diameter after anterior cervical corpectomy
SU Zai-fa,JIA Lian-shun,CHEN Xiong-sheng,et al.. A finite element analysis of comparision study on mechanical changes on cage-endplate interface between different size of the titanium mesh cages with larger or small diameter after anterior cervical corpectomy[J]. The Journal of Cervicodynia and Lumbodynia, 2011, 32(5): 323-327
Authors:SU Zai-fa  JIA Lian-shun  CHEN Xiong-sheng  et al.
Affiliation:SU Zai-fa1,JIA Lian-shun2,CHEN Xiong-sheng2,et al.(1.Department of Orthopedics,Quanzhou Hospital of TCM,Quanzhou 362000,2.Department of Orthopedics,Changzheng Hospital,The Second Military Medical University,Shanghai 200003,China)
Abstract:Objective To study the mechanical changes on cage-endplate interface between different size of the titanium mesh cages with larger or small diameter after anterior cervical corpectomy by finite element analysis.Methods Based on the intact finite element model(FE/Intact),two models were generated by simulating anterior cervical titanium mesh fusion and plate after C5 corpectomy according to the applied titanium mesh cage conditions: Φ12 mm titanium mesh cage without internal end ring(L-),Φ10 mm titanium mesh cage without internal end ring(S-),von Mises stress of seven parts on the superior and inferior cage-endplate interface was comparatively analyzed between model with L-and model with S-after loading of each model being simulated using multidirectional moment loads of 1.8 Nm combined with a compressive preload of 80 N in compression,flexion and extension.Results The stress on the superior(mean stress:6.71、7.19、9.04)Mpa and inferior cage-endplate interface(mean stress:7.61、8.23、9.97)Mpa in the model with L-was smaller than that(mean stress:7.61、8.23、9.97,and 10.65、11.59、14.26)Mpa in the model with S-under the three loading conditions.Conclusions A titanium mesh cage with larger diameter produces a significant decrease of the cage-endplate interface stress and lower incidence of titamium mesh cage subsidence after anterior cervical corpectomy.But it would be proved further if it were used in clinical practice.
Keywords:cervical spine  corpectomy  endplate  titanium mesh cage  subsidence  biomechanics  finite elememt analysis  