首页 | 本学科首页   官方微博 | 高级检索  
     


Validation of vital signs recorded via a new telecare system
Authors:Kornowski Ran  Zlochiver Sharon  Botzer Lior  Tirosh Roy  Abboud Shimon  Misan Shai
Affiliation:Department of Cardiology, Rabin Medical Center, Petach Tikva, Israel. rkornowski@clalit.org.il
Abstract:A telecare system (Medic4All) has been developed that relies on a wireless wristwatch-like sensor to measure the pulse wave from the radial artery. From this, the heart rate and respiration rate are derived. The system's performance was examined by comparing the results obtained from the pulse wave signal with those obtained from conventional electrocardiographic and spirometer devices. A total of 144 patients participated in the study; their mean (SD) age was 43 (18) years. There were 44 cardiac patients in group 1 and 100 healthy patients, who were studied in their homes, in group 2. There was a significant correlation between the heart rates measured by the two monitoring methods. A 'difference versus average' analysis showed that the error distribution had a mean (SD) value of -0.1 (3.3) beats/min. Similarly, the respiration rates measured by the two techniques were significantly correlated. The error distribution had a mean (SD) value of 0.1 (1.9) respirations/min. The present study suggests that the wrist-worn sensor represents a promising tool for online detection and monitoring of vital signs in the home.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏