Nucleus 24人工耳蜗系统直电极与弯电极神经反应遥测值的比较
作者姓名:Tao Y  Zheng Y  Wang K  Meng Z  Xu K
作者单位:四川大学华西医院耳鼻咽喉科
摘    要:目的:测试和比较Nucleus 24人工耳蜗系统中直电极与弯电极神经反应遥测(NRT)值的差别,分析内耳螺旋神经节细胞对不同电极电刺激反应特性的差异。方法:对46例进行耳蜗植入和编程的Nucleus 24人工耳蜗系统直电极与弯电极患者进行配对分析,计算患者vNRT值的均值,并进行统计学比较。结果:直电极的NRT值为160~170μV,弯电极为150~160μV,总的趋势是低频值较低而高频值较高。结论:Nucleus 24人工耳蜗系统弯电极较直电极所需电刺激量略低。

关 键 词:人工耳蜗植入  神经反应遥测  听神经复合动作电位

Compare NRT value of straight electrode and contour electrode in Nucleus 24 cochlear implant system
Tao Y,Zheng Y,Wang K,Meng Z,Xu K.Compare NRT value of straight electrode and contour electrode in Nucleus 24 cochlear implant system[J].Journal of Clinical Otorhinolaryngology,2012,26(1):8-10.
Authors:Tao Yong  Zheng Yun  Wang Kai  Meng Zhaoli  Xu Ke
Institution:Department of Otolaryngology, West China Hospital, Sichuan University, Chengdu 610041, China.
Abstract:Objective:To learn the neural response of spiral ganglion cell to electrical stimulus by comparing the neural response telemetry(NRT) value between straight electrode and contour electrode in Nucleus 24 cochlear implant system.Method:Fourty-six patients with Nucleus 24 cochlear implants were paired into two groups: 23 with contour electrode and 23 with straight electrode respectively.NRT value of each group were tested and compared.Result:The NRT value of straight electrode fell in the range of 160 μV to 170 μV and contour electrode 150 μV to 160 μV.The NRT were going up higher a little bit from electrode twenty to electrode one.Conclusion:The contour electrode requires less power consumption than straight electrode in Nucleus 24 cochlear implant system.
Keywords:cochlear implant  neural response telemetry  auditory nerve complex action potential