首页 | 本学科首页   官方微博 | 高级检索  
检索        


Interaural attenuation for tubephone insert earphones
Authors:D A Sklare  L J Denenberg
Institution:Clinical Audiology, National Institutes of Health, Bethesda, Maryland.
Abstract:Interaural attenuation of pure tone and speech signals was evaluated for a new audiometric insert earphone, the ER-3A tubephone, and a conventional TDH-49P supra-aural earphone in seven unilaterally deaf adult subjects. These results validate and extend the interaural attenuation data reported by the manufacturer of the ER-3A and his associates. At frequencies of 0.5 to 1 kHz, mean interaural attenuation for the deeply inserted ER-3A decreased from 94+ dB to 81 dB, with the lowest value for any subject, 75 dB. Interaural attenuation for speech approximated that of the 1 to 2 kHz frequency range. The ER-3A tubephone provides significantly greater acoustic isolation between the two ears in the low-mid frequency audiometric range than the conventional supra-aural earphone.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏