首页 | 本学科首页   官方微博 | 高级检索  
     


Digital radiography with photostimulable storage phosphors. Control of detector latitude in chest imaging.
Authors:H G Chotas  C E Ravin
Affiliation:Department of Radiology, Duke University Medical Center, Durham, North Carolina 27710.
Abstract:RATIONALE AND OBJECTIVES. A widely used digital radiography system based on a photostimulable storage phosphor (PSP) detector was analyzed with regard to radiographic contrast changes that result from the adjustment of detector latitude (x-ray sensitivity range) in the normal processing of chest images. METHODS. Images of an acrylic step wedge were acquired using the digital system in a mode that permitted direct control of effective detector latitude. The images were post-processed in conditions duplicating those used for portable chest examinations, and contrast was measured. RESULTS. Increases in effective detector latitude provided only marginal radiographic contrast gains in the subdiaphragm-equivalent areas of the laser-printed digital film image, while causing large reductions in radiographic contrast in the lung-equivalent region. CONCLUSION. Detector latitude is an important variable that should be monitored or controlled in investigations that compare reader performance using conventional and digital systems.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号