首页 | 本学科首页   官方微博 | 高级检索  
检索        

凝血质控品复溶后稳定性研究
引用本文:况妮娜.凝血质控品复溶后稳定性研究[J].安徽医学,2012,33(7):897-898.
作者姓名:况妮娜
作者单位:六安市第四人民医院检验科,237000
摘    要:目的通过对凝血质控品复溶后冷冻保存不同时间PT、APTT、TT的检测,探讨凝血质控品复溶后冷冻保存对其稳定性的影响。方法每周一取1瓶干粉状凝血质控品复溶后分装于5支具盖离心管中,留1支当天检测,其余4支-20℃冷冻保存,于周二至周五毎天均取1支,快速复溶后随标本一起进行质控测试,测定结果填写到质控图中。结果凝血质控品复溶后PT、APTT、TT的日间CV值均小于5%,分装后第2~5 d的测定结果与开瓶当日的测定结果间差异无统计学意义(P0.05)。结论凝血质控品复溶后-20℃冷冻保存5 d内测定结果稳定。

关 键 词:凝血  质控品  冷冻保存  稳定性

Stability of coagulation quality control materials after reconstitution
Institution:Kuang Ni’na Department of Laboratory,the Fourth People’s Hospital of Lu’an City,Lu’an 237000,China
Abstract:Objective To explore the effects of stability of coagulation control materials after reconstitution through PT,APTT,TT testing at different times after the cryopreservation.Methods A bottle of dry powder coagulation control materials was taken at Monday,reconstituted and packed in 5 cover tubes,leaving one for testing at once,and the remaining were cryopreserved at-20℃ for Tuesday to Friday testing,and the results were filled out in the quality control chart.Results CV values among day of PT,APTT,TT in coagulation control materials after reconstituted were less than 5%,and the results showed no significant difference between 2~5 days after packing and the date of the determination(P>0.05).Conclusion The measurement results of coagulation control materials are stable in 5 days of-20℃ cryopreservation after reconstituted.
Keywords:Coagulation  Quality control materials  Cryopreservation  Stability
本文献已被 CNKI 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号