首页 | 本学科首页   官方微博 | 高级检索  
     


An evaluation of tympanic thermometry in a paediatric emergency department
Authors:El-Radhi A S  Patel S
Affiliation:Queen Mary's Hospital, Sidcup, Kent DA14 6LT, UK. sahib.el-radhi@qms.nhs.uk
Abstract:

Background and objectives

The consequences of missing fever in children can be grave. Body temperature is commonly recorded at the axilla but accuracy is a problem. This study aimed to evaluate the accuracy of a tympanic thermometer in the paediatric emergency setting.

Method

In a total of 106 infants, the body temperature was measured in the daytime with an infrared tympanic thermometer, and at the axilla with an electronic thermometer and at the rectum (gold standard for measurement of body temperature). Fever was defined as a rectal temperature of 38.0 °C or greater, axillary temperature of 37.5 °C or greater, and tympanic temperature of 37.6 °C or greater. The temperature readings at the three sites were compared statistically.

Results

There was a greater correlation of the tympanic measurement with the rectal measurement than the axillary with the rectal in both febrile and afebrile infants. The mean difference between the tympanic and rectal measurements was lower than that between the axillary and rectal measurements in both groups of infants (tympanic 0.38 °C and 0.42 °C, and axillary 1.11 °C and 1.58 °C, respectively). Tympanic measurements had a sensitivity of 76% whereas axillary measurements had a sensitivity of only 24% with rectal temperatures of 38–38.9 °C.

Conclusion

Tympanic thermometry is more accurate than measurement of tempeature with an electronic axillary thermometer. It is also quick and safe, and thus it is recommended in the paediatric emergency setting.
Keywords:accident and emergency   body temperature measurement
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号