首页 | 本学科首页   官方微博 | 高级检索  
检索        


The repeat reliability of somatosensory evoked potentials
Authors:H Strenge
Institution:Abteilung Neurologie der Universit?t Kiel.
Abstract:The test-immediate-retest reliability of latency and amplitude values of cervical and cortical somatosensory evoked potentials (SEP) to median nerve stimulation was assessed in 86 normal subjects aged 15 to 71 years. In addition to the stability of data between repeat trials within one test session the standard errors of measurement and the interpretable differences for SEP measures were calculated according to measurement theory. The study revealed retest correlations rtt greater than 0.80 for all latency measures of the cervical and cortical SEPs and all cortical amplitude parameters. The highest stability was found for the latency measures of the cervical components P10, N11, N13, the cortical components P16 and N20 and for the amplitude N20/P25.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏