两种全自动血沉仪检测结果的比较分析 |
| |
引用本文: | 宫存杞,任瑞霞.两种全自动血沉仪检测结果的比较分析[J].甘肃医药,2014,33(2):100-101. |
| |
作者姓名: | 宫存杞 任瑞霞 |
| |
作者单位: | 宫存杞 (济宁市第一人民医院检验科,山东济宁,272011); 任瑞霞 (济宁市第一人民医院检验科,山东济宁,272011); |
| |
摘 要: | 目的:通过对两台不同全自动血沉分析仪检测结果的比对,实现同一实验室不同仪器间检测结果的一致性。方法:随机抽取50例标本,分别用两台仪器检测血沉,对数据进行统计学分析。结果:两台仪器血沉检测结果的相关性欠佳(r=0.92),结果差异具有统计学意义(P〈O.05)。结论:两台血沉仪检测结果不一致,应对仪器参数重新设定。
|
关 键 词: | 血沉仪 血沉 |
Comparative analysis of different automated analyzers on detecting erythrocyte sedimentation rate |
| |
Authors: | GONG Cun-qi REN Rui-xia |
| |
Institution: | (Department of Clinical Laboratory, the First People's Hospital of Jining City, Jining 272011, China) |
| |
Abstract: | Objective: To compare the consistency of two different automated analyzers in detecting erythrocyte sedimentation rate. Methods: Fifty patients were selected randomly and detected erythrocyte sedimentation rate with two different automated erythrocyte sedimentation rate analyzer and in which the data were statistically analyzed and compared. Results: The consistency of two instruments was poor (r=0.92) in which the result of the existence was significant differences (P〈0.05). Conclusion: Two blood sedimentation instrument test results are not consistency in which the instrument parameters should be reseted. |
| |
Keywords: | erythrocyte sedimentation rate analyzer |