Development of a randomised contrast detail digital phantom for observer detectability study |
| |
Authors: | Nizam Ms Ng Kh Abdullah Bjj |
| |
Affiliation: | Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia. |
| |
Abstract: | The accuracy and the efficacy of radiological diagnosis depend, to a large extent, on the conditions under which radiographs and images are viewed. This mainly involves the luminance of the display devices and the ambient room illumination. We report a perceptual study to investigate the relationship between detectability and monitor luminance as well as ambient illuminance. A statistical test pattern was used in this study, and the test pattern was developed using Microsoft® Visual Basic 6. The test pattern contained a set of randomised contrast detail objects, that is, disks of different diameters (0.7, 1.0, 1.4, and 2.0 mm) and contrasts against a black background (2.7, 3.9, 5.5, and 7.8%), simulating lesions in digital images. The receiver operating characteristic (ROC) analysis was used in this study. The results indicated that a set of optimal viewing conditions exists and that it has a significant effect on detectability performance. |
| |
Keywords: | Image quality contrast detail perception receiver operating characteristics |