首页 | 本学科首页   官方微博 | 高级检索  
检索        


Development of a randomised contrast detail digital phantom for observer detectability study
Authors:Nizam Ms  Ng Kh  Abdullah Bjj
Institution:Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia.
Abstract:The accuracy and the efficacy of radiological diagnosis depend, to a large extent, on the conditions under which radiographs and images are viewed. This mainly involves the luminance of the display devices and the ambient room illumination. We report a perceptual study to investigate the relationship between detectability and monitor luminance as well as ambient illuminance. A statistical test pattern was used in this study, and the test pattern was developed using Microsoft® Visual Basic 6. The test pattern contained a set of randomised contrast detail objects, that is, disks of different diameters (0.7, 1.0, 1.4, and 2.0 mm) and contrasts against a black background (2.7, 3.9, 5.5, and 7.8%), simulating lesions in digital images. The receiver operating characteristic (ROC) analysis was used in this study. The results indicated that a set of optimal viewing conditions exists and that it has a significant effect on detectability performance.
Keywords:Image quality  contrast detail  perception  receiver operating characteristics
本文献已被 PubMed 等数据库收录!