首页 | 本学科首页   官方微博 | 高级检索  
检索        


Tomotherapy as a tool in image-guided radiation therapy (IGRT): theoretical and technological aspects
Authors:Yartsev S  Kron T  Van Dyk J
Institution:Physics and Engineering, London Regional Cancer Program, London Health Sciences Centre, London, Ontario, Canada.
Abstract:Helical tomotherapy (HT) is a novel treatment approach that combines Intensity-Modulate Radiation Therapy (IMRT) delivery with in-built image guidance using megavoltage (MV) CT scanning. The technique utilises a 6 MV linear accelerator mounted on a CT type ring gantry. The beam is collimated to a fan beam, which is intensity modulated using a binary multileaf collimator (MLC). As the patient advances slowly through the ring gantry, the linac rotates around the patient with a leaf-opening pattern optimised to deliver a highly conformal dose distribution to the target in the helical beam trajectory. The unit also allows the acquisition of MVCT images using the same radiation source detuned to reduce its effective energy to 3.5 MV, making the dose required for imaging less than 3 cGy. This paper discusses the major features of HT and describes the advantages and disadvantages of this approach in the context of the commercial Hi-ART system.
Keywords:Image guidance  helical tomotherapy  radiation therapy  IGRT
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏