首页 | 本学科首页   官方微博 | 高级检索  
     

GE 1.5T Signa HD磁共振梯度故障实例分析
引用本文:张毅,苏根元,贾卫伟. GE 1.5T Signa HD磁共振梯度故障实例分析[J]. 中国医学装备, 2012, 0(9): 88-90
作者姓名:张毅  苏根元  贾卫伟
作者单位:中国中医科学院西苑医院放射科北京100091
摘    要:介绍GE Signa1.5T HD核磁共振工作原理,分析梯度系统的两个故障,研究排除故障的方法,总结梯度系统的维修经验,以供医学工程相关人员参考借鉴。

关 键 词:MRI  梯度系统  故障代码  维修

Analysis of breakdowns in GE 1.5T Signa HD gradient system
ZHANG Yi,SU Gen- yuan,JIA Wei-wei. Analysis of breakdowns in GE 1.5T Signa HD gradient system[J]. China Medical Equipment, 2012, 0(9): 88-90
Authors:ZHANG Yi  SU Gen- yuan  JIA Wei-wei
Affiliation:Department of Radiology, Xiyuan Hospital of China Academy of Chinese Medical Sciences, Beijing 100091, China.
Abstract:The GE Signa 1.5T HD MRI machine has dual gradient technology and new EXCITE-Ⅱ technology, which provides ultra-fast reconstruction and exclusive propeller technology so as to increase image quality and scan speed. The gradient system has complicated structure, high degree integrated system. It is difficult to repair. There are two common gradient system breakdowns. One is scan failure because of Z-axial gradient amplifier prompt over-temperature and over- current, the other is scan failure after motion artifact in zoom mode while normal scan in whole mode. This paper briefly describes the 2 breakdowns and diagnostic analysis, finds resolution and concludes repair experience. It can serve as reference for concerned personnel.
Keywords:MRI  Gradient system  Error code  Troubleshooting