First generation computerized fluoroscopic navigation in percutaneous pelvic surgery |
| |
Authors: | Mosheiff Rami Khoury Amal Weil Yoram Liebergall Meir |
| |
Affiliation: | Orthopedic Surgery Department, The Hadassah-Hebrew University Medical School, Ein Kerem, Jerusalem, Israel. ramim@cc.huji.ac.il |
| |
Abstract: | Percutaneous internal fixation of pelvic fractures has gained popularity allowing rapid mobilization with reduced surgical related morbidity; however, this method depends on conventional fluoroscopy, which exposes the patient and the surgeon to a significant amount of radiation. The use of computerized fluoroscopic navigation systems enables the simultaneous use of several radiographic projections. These preliminary fluoroscopic views are taken when the operating team stands at a distance from the radiation source. No further fluoroscopic radiation is used later during the surgical procedure. Computerized fluoroscopic navigation was used in the percutaneous insertion of 45 cannulated screws in 29 patients, including sacroiliac screws, pubic ramus screws, posterior column screws, and a supraacetabular transverse screw. Fluoroscopic verification of screw placement demonstrated a deviation
|
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|