Reduction of intra- and interlaboratory variation in CD34+ stem cell enumeration using stable test material, standard protocols and targeted training. DK34 Task Force of the European Working Group of Clinical Cell Analysis (EWGCCA)
Authors:Barnett D  Granger V  Kraan J  Whitby L  Reilly J T  Papa S  Gratama J W
Affiliation:UK NEQAS for Leucocyte Immunophenotyping, Department of Haematology, Royal Hallamshire Hospital, Sheffield, UK. d.barnett@sheffield.ac.uk
Abstract:The European Working Group on Clinical Cell Analysis (EWGCCA) has, in preparation for a multicentre peripheral blood stem cell clinical trial, developed a single-platform flow cytometric protocol for the enumeration of CD34+ stem cells. Using this protocol, stabilized blood and targeted training, the EWGCCA have attempted to standardize CD34+ stem cell enumeration across 24 clinical sites. Results were directly compared with participants in the UK National External Quality Assessment Scheme (NEQAS) for CD34+ Stem Cell Quantification that analysed the same specimens using non-standardized methods. Two bead-counting systems, Flow-Count and TruCount, were also evaluated by the EWGCCA participants during trials 2 and 3. Using Flow-Count, the intralaboratory coefficient of variation (CV) was
Keywords:CD34+ stem cell enumeration    flow cytometry    standardized protocols    clinical cell analysis
本文献已被 PubMed 等数据库收录!