引用本文:王永亮,杨建军,孙桂兰,党华. ����ﯻ�̨���Ƽ�������ֲ�����ӽ����ʺ����о�[J]. 中国实用口腔科杂志, 2010, 3(1): 30-32
作者姓名:王永亮  杨建军  孙桂兰  党华
作者单位:1.??????????????????????? ?? 266003??2.???????????????????????? ?? 266200
摘    要:目的    研制氧化锆基台并评价其与种植体连接界面的适合性。方法    2007年5月至2009年1月在青岛大学医学院附属医院口腔科选用纳米氧化锆粉,采用冷等静压成型方法和二次烧结工艺制作氧化锆基台,选取15枚氧化锆基台与0sstem USⅡ种植体装配,应用扫描电镜测量基台与种植体连接界面微间隙。结果    研制的氧化锆基台与种植体装配顺利,基台与种植体连接界面的边缘微间隙为(3.17 ± 0.34)μm、内部微间隙为(28.4 ± 2.7)μm。结论    采用纳米氧化锆粉经冷等静压成型和二次烧结工艺制作的基台与种植体有较好的适合性。

关 键 词:???????  ?????????  ???????  ?????????????????  ????  ?????  
收稿时间:2009-11-10
修稿时间:2009-12-20

The development of zirconia abutments and evaluation of the fitness of abutment-implant interface
WANG Yong-liang,YANG Jian-jun,SUN Gui-lan,DANG Hua. The development of zirconia abutments and evaluation of the fitness of abutment-implant interface[J]. chinese Journal of Practical Stomatology, 2010, 3(1): 30-32
Authors:WANG Yong-liang  YANG Jian-jun  SUN Gui-lan  DANG Hua
Affiliation:. Department of Stomatology, the Affiliated Hospital of Qingdao University Medical College, Qingdao 266003, China
Abstract:Objiective To develope a new method of manufacturing zirconia abutments and assess the fitness of abutment-implant interface. Methods The zirconia abutments were made from nano zirconia powder by cold isostatic pressing and two-step sintering. Fifteen zirconia abutments were chosen and fixed to Osstem US Ⅱ System implants, then the microgaps at abutment-implant interface were measured by scanning electron microscope. Results No difficulties were encountered in manufacturing zirconia abutments . The average width of the marginal microgaps was (3.17±0.34)microns;the width of internal microgaps at the abutment-implant interface was(28.4±2.7)microns. Conclusion The abutments made from nano zireonia powder using cold isostatic pressing and two-step sintering have good fitness with implants.
Keywords:zireonia abutments  cold isostatic pressing  two-step sintering  abutment-implant interface  mierogap  fitness