首页 | 本学科首页   官方微博 | 高级检索  
     

新生儿呼吸支持中呼气末二氧化碳分压的监测意义
引用本文:郑晨,朱莹雯,石娴静,胡壮志. 新生儿呼吸支持中呼气末二氧化碳分压的监测意义[J]. 中国小儿急救医学, 2021, 0(3): 171-175
作者姓名:郑晨  朱莹雯  石娴静  胡壮志
作者单位:苏州大学附属常熟医院常熟市第一人民医院新生儿科
摘    要:目的:比较无创面罩、经鼻导管及有创机械通气三种方法监测的新生儿呼气末二氧化碳分压(end tidal carbon dioxide pressure,PetCO 2)及其与动脉血二氧化碳分压(arterial carbon dioxide pressure,PaCO 2)的相关性。 方...

关 键 词:呼气末二氧化碳分压  动脉血二氧化碳分压  新生儿  机械通气

Monitoring significance of end-respiratory carbon dioxide partial pressure in neonatal respiratory support
Zheng Chen,Zhu Yingwen,Shi Xianjing,Hu Zhuangzhi. Monitoring significance of end-respiratory carbon dioxide partial pressure in neonatal respiratory support[J]. Chinese Pediatric Emergency Medicine, 2021, 0(3): 171-175
Authors:Zheng Chen  Zhu Yingwen  Shi Xianjing  Hu Zhuangzhi
Affiliation:(Department of Neonatology,First Hospital in Changshu City Affiliated of Soochow University,Changshu 215500,China)
Abstract:Objective To compare the neonatal end tidal carbon dioxide pressure(PetCO2)and its correlation with arterial carbon dioxide pressure(PaCO2)monitored by non-invasive mask,accessory flow nasal catheter and invasive mechanical ventilation.Methods From October 2017 to January 2020,53 cases of newborn who were needed respiratory support treatment in our hospital were selected.PetCO2was detected at admission,respiratory support and after weaning,including nasal catheter,non wound mask and invasive ventilation,and at the same time matching analysis of the corresponding with PaCO2artery blood gas analysis.Results(1)PetCO2monitored by mask was lower than PaCO2[(40.41±10.21)mmHg vs.(42.85±10.32)mmHg(1 mmHg=0.133 kPa),t=11.88,P<0.01],and there was a significant positive correlation between PetCO2and PaCO2(r=0.97,P<0.01);the mean bias of PetCO2monitored by mask was(1.20±2.31)mmHg,only 4.5%(5/110)was outside the 95%confidence interval.(2)PetCO2monitored by nasal catheter was also lower than the mean PaCO2[(40.93±10.55)mmHg vs.(42.01±10.50)mmHg,t=4.12,P<0.01],showing a significant positive correlation(r=0.96,P<0.01);the mean bias of PetCO2monitored by nasal catheter was(2.44±2.56)mmHg,and only 4.6%(7/150)was beyond the 95%confidence interval.(3)PetCO2of neonates with endotracheal intubation and mechanical ventilation was also lower than PaCO2[(43.33±10.26)mmHg vs.(49.37±11.34)mmHg,t=13.83,P<0.01],and there was also a significant positive correlation between the two groups,which was lower than that of neonates with non-invasive ventilation(r=0.94,P<0.01).The mean PetCO2bias for neonates with invasive positive pressure ventilation was(0.90±0.82)mmHg,and only 3.9%(2/51)were outside the 95%confidence interval.(4)According to gestational age,the PetCO2of early and late preterm infants was(37.25±11.32)mmHg and(39.58±10.37)mmHg,respectively,which were lower than that of full-term infants[(42.69±10.66)mmHg],and there was a positive correlation between PetCO2and PaCO2in all three groups.The correlation between PetCO2and PaCO2in early preterm infants was the lowest among the three groups(r=0.89,P<0.01).Conclusion The monitoring of PetCO2through nasal catheter,mask and invasive ventilation has a good correlation and consistency with the level of PaCO2in neonates,which can accurately reflect the level of PaCO2in neonates.The correlation between PetCO2and PaCO2in neonates with non-invasive ventilation is better than that in neonates with invasive ventilation.The correlation between PetCO2and PaCO2in late preterm infants and term infants is better than that in early preterm infants.
Keywords:End-expiratory carbon dioxide partial pressure  Arterial blood carbon dioxide partial pressure  Neonates  Mechanical ventilation
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号